CLASS II Type A2
The Class II (laminar flow) biosafety cabinet is designed to minimize the hazards inherent in working with agents assigned to biosafety levels 1, 2, 3 or 4.
Biosafety cabinets (BSCs) meet the essential requirements of EN 12469 and NSF/ANSI 49 standards for their design, construction and performance intended to protect personnel, product and the environment.
Class II (Types A1, A2., B1 and B2) biosafety cabinets are partial barrier systems that rely on air movement to provide personnel, environmental and product protection. Personnel and product protection is provided by the combination of inflow and downstream airflow captured by the cab front grille.
It has a plenum from which 30% of air is exhausted, and 70% re-circulated to the work area as the downflow.
TECHNICAL SPECİFİCATİONS
Model |
MGK-90 |
MGK-120 |
MGK-150 |
MGK-180 |
İnternal Dimensions (WxDxH) mm |
885x650x680 |
1190x650x680 |
1490x650x680 |
1790x650x680 |
External Dimensions (WxDxH) mm |
1045x800x2300 |
1350x800x2300 |
1650x870x2300 |
1960x870x2300 |
Transport Stand Dimensions (WxDxH) mm |
1045x800x890 |
1350x800x890 |
1650x870x890 |
1960x870x890 |
Average Inflow Velocity |
≥ 0,51 m/s |
|||
Average Downflow Velocity |
0,25 -0,40 m/s |
|||
Downflow Air Filter/Main Filter |
H14 HEPA, 0.3 μm particulate 99.995% (EN 1822) |
|||
Exhaust Filter |
H14 HEPA, 0.3 μm particulate 99.995% (EN 1822) |
|||
Air cleanliness class |
ISO Class 5 (ISO 14644-1)- Class 100(FED 209 E) |
|||
Bench/Working table (Standard) |
316 or 304 Stainless Steel |
|||
Front Sash/Window Standart Opening |
200mm ± 10 mm |
|||
Front Sash/Window Maximum Opening |
530mm ± 10 mm |
|||
Noise Level (Normal Mode) |
< 60 dB(A) |
|||
Noise Level (Economy Mode) |
< 55 dB(A) |
|||
llumination |
850 – 1150 Lux |
|||
Supply Voltage / Frequency |
230 V / 50 Hz |
|||
Power Consumption /Current |
1000 W / 16A |
|||
Fan |
Ebmpapst-Greentech EC |