Laminar Flow Cabinet
The Class II (laminar flow) cabinet is designed to minimize the hazards inherent in working with agents assigned to biosafety levels 1, 2, 3 or 4.
Laminar Flow cabinet meet the essential requirements of EN 12469 and NSF/ANSI 49 standards for their design, construction and performance intended to protect personnel, product and the environment.
A Laminar flow cabinet is an enclosed workstation that is used to create a contamination-free work environment through filters to capture all the particles entering the cabinet.
Laminar flow cabinet is designed to protect the work from the environment. Air moves in two way in working cabinet, one part exhausted to room at front by reventing room air move into the working zone. The rest of the internal air isrecirculated from the back holes of the cabinet by increasing HEPA filter life.
TECHNICAL SPECİFİCATİONS
Model |
MLF-90 |
MLF-120 |
MLF-150 |
MLF-180 |
İnternal Dimensions (WxDxH) mm |
931x698x651 |
1236x698x651 |
1536x698x651 |
1836x698x650 |
External Dimensions (WxDxH) mm |
995x801x2155 |
1300x801x2155 |
1600x801x2155 |
1900x801x2155 |
Transport Stand Dimensions (WxDxH) mm |
995x801x830 |
1300x801x830 |
1600x801x830 |
1900x801x830 |
Average Inflow Velocity |
≥ 0,51 m/s |
|||
Average Downflow Velocity |
0,25 -0,40 m/s |
|||
Downflow Air Filter/Main Filter |
H14 HEPA, 0.3 μm particulate 99.995% (EN 1822) |
|||
Exhaust Filter |
Not using in Laminar Flow Cabinet |
|||
Front Filter |
G4 |
|||
Air cleanliness class |
ISO Class 5 (ISO 14644-1)- Class 100(FED 209 E) |
|||
Bench/Working table (Standard) |
316 Stainless Steel |
|||
Front Sash/Window Standart Opening |
200mm ± 10 mm |
|||
Front Sash/Window Maximum Opening |
530mm ± 10 mm |
|||
Noise Level (Normal Mode) |
< 60 dB(A) |
|||
Noise Level (Economy Mode) |
< 55 dB(A) |
|||
llumination |
850 – 1150 Lux |
|||
Supply Voltage / Frequency |
230 V / 50 Hz |
|||
Power Consumption /Current |
1000 W / 16A |
|||
Fan |
Ebmpapst-Greentech EC |